2025-10-07 東京大学

ナノポア電圧マトリックス解析の概念図
<関連情報>
- https://www.s.u-tokyo.ac.jp/ja/press/10942/
- https://pubs.rsc.org/en/content/articlelanding/2025/sc/d5sc05182g
タンパク質混合物の識別のための電圧マトリックスナノポアプロファイリング Voltage-matrix nanopore profiling for the discrimination of protein mixtures
Ryo Akita, Artem Lysenko, Keith A. Boroevich, Tatsuya Yokota, Daiki Kawai, Ryo Iizuka, Tatsuhiko Tsunoda and Sotaro Uemura
Chemical Science Published:23 Sep 2025
DOI:https://doi.org/10.1039/D5SC05182G
Abstract
Solid-state nanopores are attracting attention as a label-free method for detecting diverse physical properties of biomarkers. However, improving molecular discrimination in complex biological samples remains a major challenge, partly due to uncertainty in selecting optimal measurement conditions. We developed a Voltage-Matrix Analysis that visualizes classification accuracy across multiple voltages using machine learning. We measured two tumor markers (CEA and CA15-3) individually and in mixtures using solid-state nanopores, applying Random Forest and Support Vector Machine classifiers. Overfitting occurred when baseline-involving features were used, necessitating optimization of the feature set, which led to voltage-independent high classification performance. For mixed samples, we estimated actual molecular ratios by combining classification probability histograms with detection frequency correction. We further tested mouse serum with and without centrifugation and achieved notable classification accuracy. These findings suggest that voltage-dependent structural changes influence molecular discrimination, and that our method may aid diagnosis of diseases lacking known biomarkers by identifying specific molecular population shifts.